Toshiba GA-1121 User Manual

Page of 118
A
A-10
Color Measurement Instruments
4.
Plug the small connector on the adapter cable into the side of the X-Rite DTP32 and 
plug the AC adapter into a wall outlet.
5.
Turn on the computer.
6.
Calibrate the DTP32 (see the following section).
7.
Calibrate the GA-1121 using ColorWise Pro Tools and the DTP32 (see 
).
 
Calibrating the DTP32
You will need the black-and-white X-Rite Auto-Cal Strip included with the 
densitometer. Calibrating the densitometer does not require ColorWise Pro Tools.
T
O
 
CALIBRATE
 
THE
 X-R
ITE
 DTP32
1.
Connect the densitometer to the computer and supply power (see 
2.
From the Main Menu on the X-Rite DTP32 display, press the p1 key once to reach p2.
3.
Press the cal key.
Calibrating motor speed
 is displayed, followed by the words 
INSERT CAL STRIP
.
4.
Insert the X-Rite Auto-Cal Strip into the 35mm slot on the front of the X-Rite DTP32—
arrow end first—until it stops or the densitometer starts pulling the strip.
Reading
 appears momentarily, followed by the density values and 
CALIBRATION OK
The densitometer automatically returns to the 
MAIN MENU
.
If 
UNRECOGNIZABLE STRIP
 appears, repeat the process or try cleaning the strip (see the 
X-Rite DTP32 Operating Manual).
AC adapter
Small connector