Fujifilm 4500N User Manual

Page of 129
61
4
HO
T
O
 USE 
THE OPERA
TION P
ANEL
Steps to follow
Function
Item
Calibration
Color balance
adjustment
Color matching 
On/Off selection
Brightness/contrast
adjustment
Purpose
Calibration
Color control
 
Corrects the exposure system's 
characteristics and donor 
sensitivity so that printing is 
performed under constant 
conditions.
When to adjust
 
-  When the donor has been 
replaced
-  When 1 month or more has 
elapsed since the last 
calibration
-  Before starting the color control 
adjustment
Selects the proper color matching 
function for the purpose.
Adjusts the brightness/contrast, 
at time of machine installation or 
when the monitor setting has 
been changed.
Adjusts the color balance when 
the colors on the monitor screen   
do not match those of the  
finished prints.
As a basic procedure, it is 
recommended that the 
adjustments be performed in the 
following order:  brightness,  
contrast and color balance.  
When the colors are unbalanced 
to a great extent, the adjustment 
work may be more efficiently 
performed in the following order:  
rough adjustment of brightness, 
adjustment of color balanced,  
fine adjustment of all parameters.
IMPORTANT:
Before making adjustments using the color control function, perform the calibration to make
sure that the machine is in appropriate condition.
The color control function may be used to produce prints in desired colors.  After using the
function for that purpose, be sure to restore the normal settings for the next use.
The sample image used for color matching should be one with easily identifiable transitions
between colors and gray balance.  If the sample image is biased towards particular colors, the
color adjustment will also be biased towards those colors, resulting in inappropriate color
reproduction when printing other images.
When accurate color matching is required, note the following:
It the temperature was low during the night, let the machine stand for about 1 hour after
power-on before starting printing.
If the room temperature has changed more than 5
°
C, repeat the calibration process.
4.12 
  Color Adjustment Procedures