Kodak 20R User Manual

Page of 166
Additional Calibration Information
C-22
September 2000
IMPORTANT: Whenever you click OK, the software assumes ALL changes to 
the Edit Configuration are complete and the Edit Configuration 
screen closes. Similarly, you can click Cancel at any time while 
the Edit Configuration screen is open; any changes you made to 
any of the tabbed information are canceled and the Edit 
Configuration screen closes.
Note that for Macintosh users only, you can click Default at any time to reset the 
attributes to their factory defaults and then click OK to save the defaults.
Editing Information on the Procedure Tab
The Procedure tab contains general configuration attributes as shown and 
described below. The values displayed in the example are the default settings.
  
Procedure 
Attribute
Description
Graphing
Instructs the application when to display the graphs of the data
Tolerance Level
Indicates whether the level of tolerance is Normal, Loose, or Strict
Out of Tolerance
Provides the flexibility to interact at each step or to use a fully 
automated calibration
Maximum Cycles
Indicates the maximum number of iterations the calibration 
procedure will perform if “Out of Tolerance” is set to “Automatically 
iterate up to maximum cycles.”.
Processor In Control
Indicates whether the Processor In Control dialog box is to be       
displayed at the beginning of the calibration process