Fluke 2620A User Manual

Page of 197
2620A, 2625A
Users Manual
3-20
Resistance and RTD
For all channels (0 through 20), 2-terminal resistance or RTD measurements are allowed.
Four-terminal measurements can be made on channels 1 through 10 only. Refer to
Figure 3-3.
For each channel configured for 4-terminal measurements (channels 1-10 only), a second
channel (numbered 10 higher than the first) becomes unavailable for any other type of
measurement. For example, using channel 7 to make 4-terminal resistance measurements
requires the use of input terminals for both channels 7 and 17. Channel 17 cannot be
used to take any other measurements as long as channel 7 remains configured for 4-
terminal measurements.
Table 3-14. Thermocouple Ranges
Positive Lead
Positive Lead Color
Negative Lead*
Usable Range
Type
Material
(ANSI)
(IEC)
Material
(
°
C)
J
Iron
WHITE
BLACK
Constantan
-210 to 760
K
Chromel
YELLPW
GREEN
Alumel
-270 to 1372
E
Chromel
PURPLE
VIOLET
Constantan
-270 to 1000
T
Copper
BLUE
BROWN
Constantan
-270 to 400
N
NISIL-
NOCROSIL
-270 to 1300
R
Platinum
BLACK
ORANGE
Platinum
(13% Rhodium)
0 to 1767
S
Platinum
BLACK
ORANGE
Platinum
(10% Rhodium)
0 to 1767
B
Platinum
(30% Rhodium)
GRAY
Platinum
(6% Rhodium)
0 to 1820
C**
Tungsten
WHITE
Tungsten
(26% Rhenium)
0 to 2316
* ANSI negative lead always RED, IEC negative lead always WHITE
** Hoskins Engineering Co.
Use the following procedure for resistance or RTD measurement connections to the
Input Module:
1.
 
Remove the Input Module from the rear panel.
2.
 
Loosen the two large screws on top and open the module.
3.
 
Connect wires to H (high) and L (low) terminals for each channel (channel 7 for 2-
terminal configuration or channels 7 and 17 for 4-terminal configuration, in this
example.)
4.
 
Thread these wires through the strain-relief pins and out the back of the module.
Refer to Figure 3-2.
5.
 
Close the module cover, secure the screws, and put the module back in the
instrument.