Fluke 2620A User Manual

Page of 197
2620A, 2625A
Users Manual
3-26
Table 3-17. Autoprint/Memory Storage Selection
Destination
(“dESt”) (Note
1)
Mode (“MOdE”)
(Note 2)
Press these
buttons:P
MODE
KU
)
G
D
E
G
D
E
To select from
these choices:
Print
StorE
both
ALL
ALAr
trAnS
Note 1.
“Print”
Sends data to be printed through the RS-232 interface.
“StorE”
Sends data to be stored in memory. (Hydra Data Logger only.)
“both”
Sends data to be stored and printed (Hydra Data Logger only.)
Note 2.
“ALL”
Measurements for all defined channels are printed/stored when a scan occurs.
“ALAr”
Measurements for all defined channels are printed/stored when a scan occurs
and at least one channel is in alarm.
“trAnS”
Measurements for all defined channels are printed/stored when a scan occurs
and at least one channel has transitioned into or out of alarm since the last
scan.
Table 3-18. Clearing Memory Storage
To Initiate
To Confirm
To Activate
Hold 
K
, and press 
U
“no” to exit without clearing
G
  
D
E
“CLEAr” appears in right display
“YES”
C
 to exit without clearing
Front Panel Lock out Conditions
Various methods are available to prevent accidental use of the front panel buttons. These
actions can be initiated from either the front panel or the computer interface.
Front Panel Review Only Function
Access the Review Array (press 
N, REVIEW annunciator comes on)
Press 
F and B simultaneously to activate the Review Only function (REM
annunciator comes dim).
J, H, G, and D can now be used to view the various elements of the Review Array.
All other front panel buttons are locked out, yielding a long beep when pressed. Press
both 
F and B again to deactivate the Review Only function and return the
instrument to normal front panel button operation (regular Review Array display.)
Front Panel Monitor Only Function
First place the instrument in the Monitor Function (press 
M, MON annunciator comes
on). Then press 
F and B simultaneously to activate the Monitor Only Function
(REM annunciator comes on).