Service Manual
5-18
Table 5-5. AC Volts HI Troubleshooting
Checkpoint
Signal Description
Possible Fault
A3R11 HI
Input
A3K1 through A3K14, A3U4, A3U5, A3U11, A3U12
A3L1, A3L2, A3L3
A3Z3 pin 1
Input
A3R11, A3C31, A3K15
A3U6 pin 13
Amplified (X 2.5) input
A3U7, A3Z3, A3Q3 through A3Q9, A3C15, A3C16,
A3R24, A3A25, A3R26, A3R27, A3R28, A3C23,
A3U6, A3Q13, A3U8
A3Z1 pin 2
DC equivalent of original input
A3Z1, A3U8, A3R20, A3C6, A3C7, A3C10, A3R16,
A3R17
A3U8 pin 61
DC equivalent of original input
A3Z1, A3U8, A3R20, A3C6, A3C7, A3C10, A3R16,
A3R17
5-13. Ohms Troubleshooting
Setup a channel with an open input for the desired ohms range and place the instrument
in monitor mode on that channel. Use a meter with high input impedance to measure the
open-circuit voltage at the channel input for the ohms range as listed in Table 5-6. If a
high input impedance meter is not available, only the 30-k
Ω
 and lower ranges can be
checked.
Table 5-6. Ohms Open-Circuit Voltage
Range
Voltage
300
Ω
3 k
Ω
30 k
Ω
300 k
Ω
3 M
Ω
10 M
Ω
3V
1.3V
1.3V
3V
3V
3V
If the proper voltage is not measured, setup a channel on the 300
Ω
 range (open input),
and have the instrument monitor that channel. Check for 3V dc with respect to A3TP9,
and work through the HI SOURCE and HI SENSE paths as described in Table 5-7.
If the HI path works correctly, trace continuity through the LO path. Check among A3L4
through A3L24, A3K1 through A3K14, A3R35, A3U8 pin 11, A3R43, A3K16, A3R34,
and A3U8 pin 13.
Table 5-7. Ohms HI Troubleshooting
Checkpoint
Signal Description
Possible Fault
A3U8 pin 14
Ohms Source
A3U8
A3R10 HI SRC
Ohms Source
A3R10, A3K16, A3RT1, A3Z4, A3Q10
Channel HI
Ohms Source
A3K1 through A3K14, A3U4, A3U5, A3U11, A3U12, A3L1,
A3L2, A3L3
A3U8 pin 23
Ohms Source
A3R11, A3K17, A3R42, A3C32
A3U8 pin 58
Ohms Source filter output
A3U8, A3Q2