Tektronix AFG3000 Series User Manual

Page of 108
Specifications
AFG3000 Series Quick Start User Manual
87
Modulation
AM (Amplitude Modulation)
Carrier Waveforms
Standard waveforms (except Pulse, DC, and Noise) and Arb
Modulation Source
Internal or External
Internal Modulating Waveforms
Sine, Square, Ramp, Noise, and Arb
1
Internal Modulating Frequency
2 mHz to 50.00 kHz
Depth
0.0% to 120.0%
FM (Frequency Modulation)
Carrier Waveforms
Standard waveforms (except Pulse, DC, and Noise) and Arb
Modulation Source
Internal or External
Internal Modulating Waveforms
Sine, Square, Ramp, Noise, and Arb
2
Internal Modulating Frequency
2 mHz to 50.00 kHz
Peak Deviation
DC to 5 MHz
PM (Phase Modulation)
Carrier Waveforms
Standard waveforms (except Pulse, DC, and Noise) and Arb
Modulation Source
Internal or External
Internal Modulating Waveforms
Sine, Square, Ramp, Noise, and Arb
Internal Modulating Frequency
2 mHz to 50.00 kHz
Phase Deviation Range
0.0 to 180.0 degrees
FSK (Frequency Shift Keying)
Carrier Waveforms
Standard waveforms (except Pulse, DC, and Noise) and Arb
Modulation Source
Internal or External
Internal Key Rate
2 mHz to 1.000 MHz
Numbers of Key
2
PWM (Pulse Width Modulation)
Carrier Waveforms
Pulse
Modulation Source
Internal or External
Internal Modulating Waveforms
Sine, Square, Ramp, Noise, and Arb
Internal Modulating Frequency
2 mHz to 50.00 kHz
Deviation Range
0.0% to 50.0% of pulse period
Sweep
Type
Linear or Logarithmic
Start/Stop Frequency (except Arb)
3
μHz to 10 MHz
Start/Stop Frequency (Arb)
1 mHz to 5 MHz
Sweep/Hold/Return Time
4
Range: 
1 ms to 300 s (Sweep time)
0 ms to 300 s (Hold/Return time)
Resolution: 1 ms or 4 digits
Total Sweep Time Accuracy, typical
≤ 0.4% 
1.   The maximum waveform length for Arb is 4,096. Waveform data points over 4,096 are ignored.
2.   The maximum waveform length for Arb is 2,048. Waveform data points over 2,048 are ignored.
3.   Pulse, DC, and Noise waveforms are not available. The start and stop frequencies depend on the waveform shape.
4.   Total sweep time = Sweep time + Hold time + Return time  
≤ 300 s