IBM 6300 User Manual

Page of 17
SECTION 1
INTRODUCTION
1.1 INSTRUMENT DESCRIPTION
The Model 6300 is a microprocessor controlled Visible Range Spectrophotometer covering the wave-
length range of 320 to 950nm with a 10nm bandwidth. The monochromator is of a modified Czerny Turner
design, incorporating a stepper motor driven 1200 lines/mm holographic diffraction grating and featuring
automatic second order response suppression. The 6300 has full interfacing capability for Analogue output
and serial (RS232) interfacing. The optical system is independantly housed and isolated with lenses to give
maximum protection from environmental contamination. Combined with a mechanically rigid structure, the
6300 provides a system with fast warm-up, low drift and high reliability.
1.2 INSTRUMENT SPECIFICATIONS
Wavelength:
Range
320 - 950nm
Resolution
1nm
Accuracy
±2nm
Bandwidth
10nm
Transmittance:
Range
0 to 199.9%
Resolution
0.1%
Stray Light
<0.5%
Photometric Accuracy
±2%
Absorbance:
Range
-0.300 to 1.999A
Resolution
0.001A
Concentration:
Range
-300 to 1999 Concentration
Resolution
0.1/1
Units
ppm, mg/l
-1
, g/l
-1
, M, blank
Factor
0 to 199.9, 1000 to 9999
Photometric Noise Levels:
<1%
Photometric Stability:
1%/Hr after warm-up
Readout:
Custom LCD Graphics display
Outputs:
Analogue (0 - 1.999V d.c.) / RS232 serial port
Light Source:
Tungsten Halogen 20W 12V
Input Voltage:
115/230Vac -20% + 10%
Input Power:
<50W
Size:
365 (w) x 272 (d) x 160 (h)mm
Weight:
6Kgs
1
6300/REV B/12-96