Powerware ST-2400S User Manual

Page of 17
dynamometer.  The calibration torque produced is equal to the sum of the torque 
produced by the Torque Test Bar itself plus the torque produced by the test 
weight(s) – which is equal to the total weight of the test weight(s) times the 
horizontal distance from the center line of the dyno to the position of the test 
weights.  For example, assume the torque of the torque test bar itself is 17 Ft.-
Lbs., the total weight is 100 Lbs. and the horizontal distance is 4.00 Ft.  The 
calibration torque value produced by this setup is equal to 417 Ft.-Lbs. ( = 4.00 
Ft. * 100 Lbs. + 17 Ft.-Lbs.) 
 
2.)  Shunt Resistor: The 2400S horsepower computer has an internal shunt resistor 
which can be used to check the torque span calibration. The 2400S can 
(optionally) be equipped with a precision shunt resistor which can be used to 
perform a NIST (National Institute of Standards and Time) traceable torque 
calibration.  The “torque” introduced by the shunt resistor depends on the 
resistance value of the shunt resistor as well as the properties of the load cell.  For 
the NIST traceable torque calibration to be accurate, both the shunt resistor value 
and the load cell must have NIST traceable calibrations which should be checked 
periodically.  A 2400S horsepower computer without the precision shunt resistor 
is equipped with a 1% precision 45.3 K Ohm shunt resistor (not adequate for 
NIST traceable calibrations) which can be used for a quick and fairly accurate 
check on the torque calibration and the torque sensor circuits of the 2400S.   After 
performing the calibration of the unit using the dead weight test fixture calibration 
method, remove the dead weight test fixture and set the Shunt Resistor Switch to 
ON.  The shunt resistor will produce a torque reading that is (roughly) 2/3 of the 
full scale torque of the unit (i.e. about 3300 Ft.-Lbs. on a 5000 Ft.-Lbs. full scale 
2400S).  More precise values are shown in the table below.  The reading should 
be within +/- 2 or 3 % of the appropriate value from the table below. The 
tolerance derives from the +/- 1% tolerance of the 45.3 K Ohm shunt resistor plus 
the tolerance of the load cell (likely +/- 1 or 2 %).  If the reading is not this close, 
you should double check the calibration and your procedure as a likely problem is 
indicated.  If the reading is within the above limits, record its value.  It can be 
used later for a quick check on your unit’s calibration as it should be stable over 
time and temperature to within +/- 0.25% or better. 
 
Full scale Torque (Ft.-Lbs.) 
45.3 K Ohm Shunt Resistor Torque 
100 64.14 
200 128.3 
500 320.7 
1,000 641.4 
2,000 1283 
5,000 3207 
10,000 6414 
 
 
14