B&K 0-3A User Manual

Page of 58
 
15 
 
OPERATING INSTRUCTIONS 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 5. Setting Current Limit. 
 
5.  Increase the  VOLTAGE setting until the Voltage LED display 
reads the desired value. The  FINE control permits easier setting 
to a specific value. 
6.  Note the load current on the Current LED display. 
7.  If the load current exceeds the preset current limit, the  CV 
indicator will go off and the CC indicator will light. In this case, 
the power supply automatically switches to the constant current 
mode, and further rotation of the  VOLTAGE  control will not 
increase the output voltage. 
 
SETTING CURRENT LIMIT 
 
1. Determine the maximum safe current for the device to be 
powered. 
2. Temporarily short the (+) and (-) terminals of the power 
supply together with a test lead. 
3. Rotate the  Coarse VOLTAGE  control away from zero 
sufficiently for the CC indicator to light. 
4. Adjust the Coarse and  Fine CURRENT  control for the desired 
current limit. Read the current value on the Current LED 
display. 
5. The current limit (overload protection) has now been preset. Do 
not change the CURRENT  controls settings after this step. 
6. Remove the short between the (+) and (-) terminals and hook 
up for constant voltage operation. 
 
Midrange 
 
 
 
Adjust to 
desired 
current limit  
 
 
 
CC 
 
Read current limit  
Value from meter 
Temporarily short  
(+) to ( -)