Princeton 4300 User Manual

Page of 112
20 
PI-SCX System Manual 
Version 2.E 
 
20 
#  Feature 
12.
  Shutter Setting Selector: 
Sets the shutter hold voltage. Contact factory for the correct 
shutter setting for a remote x-ray shutter.
 
13.
  Remote Shutter Connector:
 Provides shutter-drive and hold voltages for a Princeton 
Instruments 25 mm external shutter (typically, an entrance slit shutter). May be used for a remote 
x-ray shutter. 
 
WARNING:
 
Dangerous live potentials are present at the 
Remote Shutter Power connector. 
To avoid shock hazard, the Controller power should be OFF when connecting or 
disconnecting a remote shutter.
 
14.
  Power Input Module: 
Contains the power cord socket and two fuses. Depending on the 
ST-133 version, the power switch may be located directly above the power module.
 
15.
  Fuse/Voltage Label:
 Displays the controller’s power and fuse requirements. This label may 
appear below the power module.
 
16.
  USB 2.0 Connector:
 Provides two-way serial communication between the controller and the 
host computer. Uses USB 2.0 protocol.
 
 
Figure 4.  NOT SCAN and SHUTTER Signals
 
Cables 
 
 
 
 
 
 
 
Detector-Controller: 1 MHz or 100kHz/1MHz systems. The standard 10' 
cable (6050-0484) has DB-25 Male connectors with slide-latch locking 
hardware. This cable interconnects the Detector connector on the rear of the 
ST-133 with the 25-pin D connector on the back of the PI-SCX camera. The 
Detector-Controller cable is also available in 6', 15', 20', and 30' lengths. Note 
that a longer cable may degrade camera performance. 
Computer Interface Cable: Depending on the system configuration, either a 
USB or a TAXI cable will be shipped.