System Requirements 
Environmental
 
Operating temperature: 0ºC to +30ºC;  
Operating temperature range over which system specifications can be met: +18ºC to 
+23ºC 
Relative humidity <50% noncondensing.
 
Ventilation 
Camera: Allow at least one inch of clearance for side and rear air vents. 
ST-133:  There is an internal fan located at the right side of the rear panel behind an 
exhaust opening.  Its purpose is simply to cool the controller electronics. This fan 
runs continuously whenever the controller is powered. Air enters the unit through 
ventilation openings on the side panels, flows past the warm electronic 
components as it rises, and is drawn out the rear of the controller by the fan. It is 
important that there be an adequate airflow for proper functioning. As long as 
both the controller’s intake ventilation openings and the fan exhaust opening 
aren’t obstructed, the controller will remain quite cool. 
Vacuum 
The CCD is housed in the vacuum with the fiberoptic taper extending beyond the vacuum 
chamber with a thin Beryllium window to block visible light and to admit X-rays. The 
camera is shipped with a vacuum quality superior to the minimum required to assure 
proper operation of the system for an extended period of time without concern for the 
quality of the vacuum. See Appendix C for additional information.  
Coolant 
COOLANT IS HARMFUL IF SWALLOWED.                                                            
KEEP OUT OF REACH OF CHILDREN. 
PI-SCX cameras with water-assisted cooling or water-only cooling require circulating 
coolant (50:50 mixture of ethylene glycol and water) for proper operation. All hose 
connections should be secured with good quality hose clamps.  
Flow Rate:
 Users are advised to install a flow meter to monitor the rate. 
PI-SCX:1300
 
> 1 liter/minute. 
PI-SCX:4096
 
3 liters/minute (minimum). 
PI-SCX:4300
 
3 liters/minute (minimum).  
Fluid Pressure:
 25 psig (maximum).  
Inlet/Outlet Port Locations: 
PI-SCX:1300    
Two quick-disconnect coolant ports are supplied with a water-
cooled system and are mounted at the sides of the camera. For the best cooling, make 
the hose connections to the Inlet and Outlet ports as shown in the outline drawing on 
page 38. Use 1/4" thin-wall tubing. 