Princeton 4300 User Manual

Page of 112
Chapter 5 
Operation 
45 
 
 
 
45 
Experiment Setup Timing tab (Acquisition|Experiment Setup…): 
 
Timing Mode: Free Run 
 
Shutter Control: Normal 
 
Safe Mode vs. Fast Mode: Fast Mode 
General tab (Display|Layout…): Select Horizontal and Vertical Cross 
Sections. 
Acquiring Data 
If you are using WinView/32 and the computer monitor for focusing, select 
Focus
 from 
the 
Acquisition menu. Successive images will be sent to the monitor as quickly as they 
are acquired. Since no X-ray source is being used, the acquired images will be of the 
camera's dark charge. As the CCD temperature decreases, the image will get darker as the 
dark charge decreases. Signal changes may be more easily seen on the cross-section 
views. 
Figure 17 shows the kind of image data you might see in WinView.  The horizontal and 
vertical cross sections have been turned on via the 
General
 tab card (
Display|Layout
). 
 
Figure 17.  Example of WinView Data Acquired from First Light Procedure  
Because the time to acquire and read out an image varies directly with the size of the 
CCD, the observed frame rate will vary greatly depending on the CCD installed. With a 
short exposure time, it is not uncommon for the frame readout time to be significantly 
longer than the exposure time: for a Fairchild 4096x4096 array, a full frame readout at 
1 MHz takes approximately 17 seconds. 
This completes First Light. If the system functioned as described, you can be reasonably sure 
it has arrived in good working order. In addition, you should have a basic understanding of 
how the system hardware is used. A recommended procedure for powering down the camera 
is provided in the next section. Other topics, which could be quite important in certain 
situations, are discussed in the following chapters. See the appropriate application software 
manual for information on using the software to control the system.