Princeton 4300 User Manual

Page of 112
52 
PI-SCX System Manual 
Version 2.E 
 
52 
This technique is also useful in high light level experiments, where the camera is again 
photon shot-noise limited. Summing multiple pixels in software corresponds to collecting 
more photons, and results in a better S/N ratio in the measurement. 
Output Amplifier Selection 
The output amplifier amplifies the collected charge from the output node and outputs it as 
electrons/count. Although Figure 18 shows an array with a single output node and 
amplifier, some PI-SCX systems are available with dual output nodes and amplifiers (one 
set at each end of the shift register).  If your system has dual output amplifiers, you can 
choose the output amplifier to be used (High Capacity or Low Noise) via WinView/32 or 
WinSpec/32 on the 
Acquisition|Experiment Setup…|Main 
tab:  
 
High Capacity amplifier:
 Provides a spectrometric well capacity that is 
approximately 3 times the well capacity for the Low Noise amplifier selection. 
High Capacity is suitable when you have intense light signals or signals with 
high dynamic range.  
 
Low Noise amplifier:
 Provides the highest sensitivity performance and is 
suitable when you have weak signals. 
Note: 
The choice of output amplifier and analog gain setting should be considered 
together for the best signal capture.
 
Analog Gain Control 
Note:  A gain control switch is located on the rear of the PI-SCX:1300 camera but has 
been deactivated. However, analog gain is changeable via the WinView/WinSpec 
software (on the 
Acquisition| Experiment Setup…|ADC
 
tab card).  If your camera 
is not designed for analog gain selection, these settings will not be accessible in the 
software.  
Analog gain control (a function of the preamplifier) is used to change the relationship 
between the number of electrons acquired on the CCD and the Analog-to-Digital Units 
(ADUs or counts) generated.  In WinView/32 and WinSpec/32, the analog gain choices 
vary depending on the CCD array and the number of output amplifiers:   
  Single Amplifier:
 Typically, three settings are available: Low 2x, Medium 1x, 
and High 1/2x.  
  Dual Amplifier:
 Typically, three settings are available: Low 2x, Medium 1x, and 
High 1/2x (Low-noise mode) or Low 4x, Medium 2x, and High 1x (High-
capacity mode). 
The analog gain of the detector should generally be set so that the overall noise is ~1 
count RMS. In most instances this will occur with the switch set to Medium. If the array 
is a 1340 × 1300 configured with the low-noise outputLow will probably be a more 
suitable Analog Gain setting. In situations where the A/D range exceeds that of the array, 
it will generally be better to set the Analog Gain to High so that the signal can be spread 
over as much of the A/D range as possible. This is a particularly important consideration 
in absorbance measurements. Users who consistently measure low-level signals may 
wish to select High, which reduces some sources of noise. Users who measure high-level 
signals may wish to select Low to allow digitization of larger signals. Customized values 
of gain can be provided. Contact Princeton Instruments for additional information.