Princeton 4300 User Manual

Page of 112
 
55 
 
Chapter 6 
Advanced Topics 
Introduction 
Previous chapters have discussed setting up the 
hardware and the software for basic operation. 
This chapter discusses topics associated with 
experiment synchronization (set up on the 
Experiment Setup|Timing
 tab in WinView and 
WinSpec).With the exception of Edge Trigger, the 
topics are addressed in order of their appearance 
on the Timing tab (see Figure 21).  
"Timing Modes"
, the first topic, discusses 
standard Timing Modes, Shutter Control, and 
Edge Trigger. Also included under this topic is a 
discussion of the 
EXT SYNC
 connector, the input 
connector for a trigger pulse. 
"Fast Mode and Safe Mode"
, the second topic, 
discusses the Fast and the Safe speed modes. Fast 
is used for real-time data acquisition and Safe is 
used when coordinating acquisition with external 
devices or when the computer speed is not fast 
 
Figure 21.  Timing tab 
enough to keep pace with the acquisition rate. 
"TTL Control"
, the final topic, discusses the TTL IN/Out connector on the rear of the 
ST-133 and how commands can be sent via TTL levels to and from the ST-133. 
Timing Modes 
The chart to the right lists the timing mode 
combinations (selected on the
 
Experiment 
Setup|Timing 
tab). Use this chart in 
combination with the detailed descriptions in 
this chapter to determine the optimal timing 
configuration.  
The basic timing modes are Free Run, 
External Sync, and External Sync with 
Continuous Cleans.  These modes are 
combined with the Shutter options to provide 
the widest variety of timing modes for 
precision experiment synchronization. 
Mode 
Shutter 
Free Run 
Normal 
External Sync 
Normal 
External Sync 
PreOpen 
External Sync with 
Continuous Cleans 
Normal 
External Sync with 
Continuous Cleans 
PreOpen 
Table 3.  Camera Timing  Modes 
The shutter options available include Normal, PreOpen, Disable Opened or Disable 
Closed. Disable simply means that the shutter will not operate during the experiment. 
Disable closed is useful for making dark charge measurements or when no shutter is 
present. PreOpen, available in the External Sync and External Sync with Continuous