Princeton 4300 User Manual

Page of 112
70 
PI-SCX System Manual 
Version 2.E 
 
70 
deterioration in new cameras may be faster than in old ones. For example, a camera that 
has to be repumped after perhaps a year of operation, may not have to be pumped again 
for several years. In any case you may notice a gradual deterioration in temperature 
control performance indicative of vacuum deterioration. If this happens, the camera will 
have to be repumped. Appendix C provides directions for repumping the vacuum. If you 
have the appropriate equipment and personnel with the necessary expertise available, you 
may wish to pump down the camera at your facility by following the procedure outlined 
in Appendix C. If the necessary equipment and expertise isn’t available, simply contact 
the factory to make arrangements for returning the camera to the support facility nearest 
to you to have the vacuum restored.  
Data Loss or Serial Violation 
You may experience either or both of these conditions if the host computer has been set 
up with Power Saving features enabled. This is particularly true for power saving with 
regard to the hard drive. Make sure that Power Saving features are disabled while you are 
running WinView/32. 
Data Overrun Due to Hardware Conflict message 
  
Figure 34.  Data Overrun Due to Hardware Conflict dialog 
If this dialog appears when you try to acquire a test image, acquire data, or run in focus 
mode, check the CCD array size and then check the DMA buffer size. A large array (for 
example, a 2048x2048 array), requires a larger DMA buffer larger setting than that for a 
smaller array (for example, a 512x512 array). 
To change the DMA buffer setting: 
1.  Note the array size (on the 
Setup|Hardware|Controller/CCD
 tab or the 
Acquisition|Experiment Setup|Main
 tab Full Chip dimensions). 
2.  Open 
Setup|Environment|Environment
 dialog. 
3.  Increase the DMA buffer size to a minimum of 32 Mb (64 Mb if it is currently 
32 Mb or 128 Mb if it is currently 64 Mb), click on 
OK
, and close the WinX 
application. 
4.  Reboot your computer. 
5.  Restart the WinX application and begin acquiring data or focusing. If you see 
the message again, increase the DMA buffer size.