Princeton 4300 User Manual

Page of 112
Chapter 7 
Troubleshooting 
83 
 
 
83 
Serial violations have occurred. Check interface cable. 
 
Figure 46.  Serial Violations Have Occurred dialog 
This error message dialog will appear if you try to acquire an image or focus the camera 
and either (or both) of the following conditions exists:  
 
The camera system is not turned ON.  
 
There is no communication between the camera and the host computer. 
To correct the problem: 
1.  Turn OFF the camera system (if it is not already OFF).  
2.  Make sure the Detector-Controller cable is secured at both ends and that the 
computer interface cable is secured at both ends. 
3.  After making sure that the cables are connected, turn the camera system power 
ON
4.  Click 
OK
 on the error message dialog and retry acquiring an image or running in 
focus mode.  
Note: This error message will also be displayed if you turn the camera system OFF or a 
cable comes loose while the application software is running in Focus mode.