Appendix C 
Repumping the Vacuum 
Introduction 
Fiberoptic taper CCD cameras are shipped with a vacuum level below 10
-5
 Torr to assure 
proper cooling performance and to prevent condensation from collecting on the CCD. In 
time, the vacuum level could deteriorate to where achieving temperature lock will no 
longer be possible. If this happens, it will be necessary to restore the vacuum. 
A ¼
 Vacuum Pumping fitting (Figure 56) is supplied with the camera but is not factory 
installed. The access port, located at the top of the camera, is capped. The cap can easily 
be removed, allowing the vacuum pumping fitting to be installed.  
 
Figure 56.  Vacuum Pumping Fitting  
A special Indium-tipped Vacuum Seal (Figure 57) is installed in the side of the camera 
(see Figure 15 on page 39). At installation, the Vacuum Seal’s Indium tip is crushed to 
provide a metal-to-metal seal in the vacuum chamber. The number of times this can be 
done before a new Vacuum Seal is required is limited. Typically, an Indium tip can be 
crushed up to ten times before the Vacuum Seal must be replaced to assure a good 
vacuum seal. Because the tip is crushed only once at the factory, it can be successfully 
used again numerous times in the field. With repeated repumpings, however, it will 
eventually be necessary to replace the Indium-tipped Vacuum Seal (PI #2550-0352). 
Note: The required seating torque for successful vacuum sealing is 28 in. lb. (3.2 Nm). 
 
Figure 57.  Indium-tipped Vacuum Valve