Kodak i800 Series User Manual

Page of 83
 
 
A-61169 July 2003 
6-5 
Message Listing 
Following is a list of messages and corrective actions you can take if one 
of the following messages appears on the Operator Control Panel. Most 
corrective actions can be found in this User’s Guide; appropriate chapters 
are referenced. 
If one of the following messages appears, take the following action: 
• 
Determine whether or not the error can be addressed from the host 
system. 
• 
If the error cannot be addressed by the host system, attempt to clear 
the error by taking the action listed below. 
• 
If the prescribed action(s) does not clear the error, initiate a controlled 
power-down sequence: 
Controlled Power-down Sequence 
1.  Wait for all information currently held in the buffer to be transferred to 
the host system. 
2.  Turn off the scanner. 
3.  Wait 30 seconds. 
4.  Turn the scanner back on. 
If the error condition is not cleared after using the controlled power-down 
sequence, call Service. 
Message 
Description 
Message 
Action 
Calibrate scanner now 
You must calibrate the scanner now. See Chapter 3, Operator Control Panel 
Functions
 for calibration procedures. 
Message 
Action 
Calibrate scanner soon 
Calibrate the scanner when the current batch has completed. See Chapter 3, 
Operator Control Panel Functions for calibration procedures. 
Message 
Action 
Calibration failed 
Perform a controlled power-down sequence and recalibrate the scanner. Be sure 
you are using the correct calibration target. Clean the imaging guides. See 
Chapter 3, Operator Control Panel Functions for the correct calibration target. If 
the message persists, call Service. 
Message 
Action 
Calibration succeeded 
Informational message. The scanner has been successfully calibrated.