5-2
A-61371   September 2004 
Carrier Glass (option)
NOTE: Before cleaning the carrier glass, remove the projection lens 
from the scanner. Refere to Chapter 3, “Installing the projection 
lens” for procedures.
• With a damp cloth, clean and remove any dust or debris from the 
surface of the carrier glass. 
To open the carrier glass:
• Pull the handle of the microfiche holder. With a damp cloth, clean and 
remove dust and debris by wiping the inner surfaces of the carrier 
glass.
Replacing the 
projection lamp
Use the following procedure to replace the projection lamp whenever a 
reduction in brightness on the screen is detected or whenever the lamp 
burns out. Make sure that the replacement lamp is specified for use 
with this scanner.
If the projection lamp burns out during a print operation, an L2 code 
appears and the print job will stop (a blank piece of paper may be 
output depending upon the stage of the job).
Once the system detects that the projection lamp has burnt out, the 
scan job will be cancelled. If the system is operating in the Cycle Print 
mode, the scanner operation will stop.
To replace the projection lamp:
1. Turn the power switch off and unplug the power cord from the wall 
socket-outlet. Grasp the power cord plug only when unplugging the 
power cord.
WARNING:Never touch the lamp with bare hands just after 
operation or immediately after turning it off. Wait 
at least 10 minutes after shutting the power off to 
allow the lamp to cool down as it becomes very 
hot during use.