Honeywell SC1000 User Manual

Page of 174
 
SC Series Instruction Manual     page 117
Hig
h-L
ev
el
 In
pu
t C
han
nel
12
12.6.6 
CALIBRATE
 Menu 
Item
This menu item performs a calibration according to what was entered in the 
CALI-
BRATION TYPE
 and 
CALIBRATION DATA
 menu items.
If the 
CALIBRATION TYPE
 is Shunt Calibration...
The display will read 
DOING SHUNT CAL
, and prompt you to 
APPLY 00000 UNIT
 (where 
00000 and UNIT are previously entered as
 ZERO SCALE VALUE
 and 
DISPLAY. UNITS
When you have applied this load to the transducer, press [ENTER].   
The display will then read 
WORKING
 and calibration will take place.  Upon com-
pletion, the display will indicate 
DONE
 and the instrument will return to the RUN 
mode.
If the 
CALIBRATION TYPE
 is 2-Point Known Load Calibration...
The display will read 
DOING 2POINT CAL
, and prompt you to 
APPLY 00000 UNIT
 
(where 00000 and UNIT are previously entered as
 KNOWN POINT 1/2
 and 
DISPLAY. 
UNITS
). When you have applied this load to the transducer, press [ENTER]. 
The display will read 
WORKING
, then 
APPLY 01000 UNIT
 (where 01000 and UNIT are 
previously entered as
 KNOWN POINT 2/2
 and 
DISPLAY. UNITS
). When you have 
applied this load to the transducer, press [ENTER].
Upon completion, the display will indicate 
DONE
 and the instrument will return to 
the RUN mode.
If the 
CALIBRATION TYPE
 is 3-Point Known Load Calibration...
The display will read 
DOING 3POINT CAL
, and prompt you to 
APPLY 00000 UNIT
 
(where 00000 and UNIT are previously entered as
 KNOWN POINT 1/3
 and 
DISPLAY. 
UNITS
). When you have applied this load to the transducer, press [ENTER]. 
The display will read 
WORKING
, then 
APPLY 00500 UNIT
 (where 00500 and UNIT are 
previously entered as
 KNOWN POINT 2/3
 and 
DISPLAY. UNITS
). When you have 
applied this load to the transducer, press [ENTER]. 
The display will read 
WORKING
, then 
APPLY 01000 UNIT
 (where 01000 and UNIT are 
previously entered as
 KNOWN POINT 3/3
 and 
DISPLAY. UNITS
). When you have 
applied this load to the transducer, press [ENTER]. 
Upon completion, the display will indicate 
DONE
 and the instrument will return to 
the RUN mode.
NOTICE
Before performing a calibration,
the transducer must be connected to the instrument, 
the channel’s Excitation and Signal Jumpers must be set, 
the 
CALIBRATION TYPE
the 
CALIBRATION DATA
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.