Fuji Labs ZSP User Manual

Page of 62
– 9 –
7.3
General operation of indication/operation panel
The measuring mode is assumed when
power is turned ON.  The gas concentra-
tion appears on the main indication, while
the range being used appears on the sub in-
dication.
Advice on Operation
✩When selecting the range
In the setting status, as shown in the figure
at the right, the high range is selected when
 key is pressed, while the low range is
selected when   key is pressed.
✩When selecting the gas component
In the setting status, as shown in the figure
at the right, the gas component can be set
by pressing  COMP  key.
Example: When  COMP  key is pressed
while the 1st component is flash-
ing, the flashing moves to the
2nd component (figure at right).
The setting for the 2nd compo-
nent is now changeable.
* The 1 component type analyzer is not pro-
vided with  COMP  key.
✩For releasing zero or span
If a mistake has been made in zero or span
calibration, then perform a reset in the fol-
lowing way.
The figure at the right shows an example.
If  SPAN  key has been pressed mistakenly
instead of  ZERO  key, then press  SPAN
key again.  Calibration is released.
If   ZERO  key has been pressed mistak-
enly instead of  SPAN   key, the press
ZERO  key again.
Calibration is released.
Gas concentration
indication
Range
   indication
NO
Hige rang is selected 
when      key is pressed
Low rang is selected 
when      key is pressed
1st  component
2rd  component
3rd  component
COMP
ENT ZERO SPAN CAL
ZERO SPAN
ZERO SPAN
ZERO SPAN
Pressing changes the settable gas component.  
Pressed wrongly.
The lamp blinks.
Press again to extinguish the lamp.
Thus reset. Press the              key afresh.
ZERO