Fuji Labs ZSP User Manual

Page of 62
– 12 –
8.3
Calibration concentration setting (except zirconia O
2
 sensor)
Set a span value for the calibration concentration setting. (Zero point calibration concentration is fixed at
zero.)
Set the calibration gas concentration (span value).
When  FUNC  key is pressed in the measuring sta-
tus, the previously set span value will appear on
the main indication.
The CAL SET LED of the function indicating
lamps will flicker.
By pressing the 
 key, the highest digit of the
main indication for the 1st component will flicker,
and the span value can not be set.
Select a range with the     keys.
After selecting the range, set a span value in this
status.
The numeric value will be incremented by press-
ing 
 key.
The digit to be set can be selected by pressing 
key.
For selecting the 2nd and subsequent compo-
nents and the range, press the 
 key and
while the main indication for the 1st compo-
nent is flickering, press the  COMP  key and
then the main indication for the component to
be set will flicker.  Now press the     keys to
select the range to be set.
When the span value has been set, press the  ENT
key.  Setting operation is completed.
ENT
CAL  SET
FUNC
NO
NO
NO
NO
NO