Fuji Labs ZSP User Manual

Page of 62
– 17 –
8.6
Span calibration (low concentration point calibration in case of zirconia O
2
 sensor)
Carry out span calibration by flowing calibration gas at
the concentration set as a span value.
Flow the calibration gas (0.3 N/min.) by adjusting the
needle valve of pressure regulator.  Wait until the indica-
tion stabilizes, and then conduct span calibration.
Press the  SPAN  key and the  SPAN  key lamp flickers.
In the case of a multi-component analyzer, the main
indication will flicker for the components which
can be calibrated.
Press the COMP key and the settable gas compo-
nent will change.
When the gas component has been determined,
press the     keys and select a range.
Note: The lamp will not flicker in the case of a
single-component analyzer.
Next press the  CAL  key and conduct span calibration.
The  CAL  key lamp lights steadily during calibration.
Since the measuring status is resumed after calibration,
continue the calibration for each component and each
range.
Note: On the low range of zirconia type O
2
 sensor, when zero gas is calibrated with 9 to 10 Vol. % O
2
,
the zero and span calibrations interfere with each other.  So, each calibration should be re-
peated a few times.
When the function for automatic calibration is provided, the solenoid valve turns on and the
span gas for the selected component flows upon pressing the SPAN key.
NO
NO
NO
CAL
NO
SPAN
(                  )
CAL
SPAN
ZERO
ENT
Flickers
CAL
SPAN
ZERO
ENT
Flickers
COMP
CAL
SPAN
ZERO
ENT
lights
CAL
SPAN
ZERO
ENT
(In case of a multi component analyzer)
( Lamp is turned off after calibration )