Fuji Labs ZSP User Manual

Page of 62
– 45 –
14.  SPECIFICATIONS
Specifications
Type of cubicle:
Outdoor, self-standing, sealed type or indoor type
Measuring system:
Non-dispersion infrared absorption (NDIR method)  for NO
x
, SO
2
, CO and CO
2
;
zirconia method for O
2
.
Sampling system:
Dry sampling electronic cooling & dehumidification system
(Permapure dryer adopted)
Measuring range:
NO
x
— 0 to 100 .... 2000 ppm
SO
2
— 0 to 100 .... 1000 ppm
CO — 0 to 200 .... 2000 ppm
CO
2
— 0 to 5 ........ 50%
O
2
— 0 to 10/25%
Note: Measuring range of each gas is 50 to 5000ppm, available in
approved type (special order).
Repeatability:
±
0.5% of full scale
Drift:
Zero —
±
2% of full scale/week for NOx, SO
2
 and CO
±
2% of full scale/month for O
2
Span —
±
2% of full scale/week
Linearity:
±
2% of full scale
Response time:
90% response from the inlet of cubicle
SO
2
 : 4 min. CO : 2.5 min NO : 2 min
Others : 3 min.
Output:
4 to 20mA DC (allowable load resistance 550
Ω
), non-isolated.
For the number of output points, see Table 1.
Contact output:
Each NO (1a) contact (100V AC, 1A) for error (analyzer error, calibration error) and
auto calibration status
Indication:
Digital indication (indicating measured value and measuring range on analyzer front
panel of infrared gas analyzer)
Sample gas extracting rate: Approx.   2 N/min.
Gas extractor:
Electrical heating type (filter built in)
Filter mesh — 40
µ
m mesh of 316 stainless steel
Probe
— 316 stainless steel, length 600, 800 or 1000mm
Flange
— JIS 5 K65AFF
Sample inlet tube:
ø 10/ ø 8mm teflon tube for CO and O
2
 meter and NO
x
 and O
2
 meter
Heating tube ( ø 10/ ø 8mm teflon tube) for NO
x
, SO
2
 and O
2
 meter and NO
x
, SO
2
, CO
and O
2
 meter
Standard gas (3.4 N):
Standard air gas for zero calibration and O
2
 span calibration
Standard analyte/remaining N
2
 gas for low range span calibration and O
2
 zero calibration
Form verification:
Form for all components already approved
Ambient temperature:
-5 to +40
°
C
Power supply:
100
±
10V AC, 50 or 60Hz
Power consumption:
Approx. 1,200 VA (excluding consent and heating tube)
Mass (weight):
Approx. 300 kg (excluding standard gas)
Finish color:
Munsell 5Y 7/1 semigloss
(Range selection:
Max. rate 1:20
(except for O
2
))