Tektronix TDS 700D User Manual

Page of 154
Performance Tests
1–26
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications 
TDS 794D Test Record
Photocopy this and the next two pages and use them to record the performance
test results for your TDS 794D.
TDS 794D Test Record 
Instrument Serial Number:
Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
TDS 794D Performance Test
Minimum
Incoming
Outgoing
Maximum
Offset Accuracy
CH1 Offset
+10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
CH2 Offset
+10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
CH3 Offset
+10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
CH4 Offset
+10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
DC Voltage Measurement Accuracy (Averaged)
CH1 
10 mV Vert scale setting, 
–5 Div position setting, +.5 V offset
+0.5755 V
__________
__________
+0.5845 V
CH1 
10 mV Vert scale setting, 
+5 Div position setting, –.5 V offset
–0.5845 V
__________
__________
–0.5755 V
CH1
200 mV Vert scale setting, 
0 Div position setting, +5 V offset
+5.5345 V
__________
__________
+5.6655 V
CH1
200 mV Vert scale setting, 
0 Div position setting, –5 V offset
–5.6655 V
__________
__________
–5.5345 V
CH1
1 V Vert scale setting, 
0 Div position setting, +2.5 V offset
+5.289 V
__________
__________
+5.711 V
CH1
1 V Vert scale setting, 
0 Div position setting, –2.5 V offset
–5.711 V
__________
__________
–5.289 V
CH2 
10 mV Vert scale setting, 
–5 Div position setting, +.5 V offset
+0.5755 V
__________
__________
+0.5845 V
CH2 
10 mV Vert scale setting, 
+5 Div position setting, –.5 V offset
–0.5845 V
__________
__________
–0.5755 V
CH2
200 mV Vert scale setting, 
0 Div position setting, +5 V offset
+5.5345 V
__________
__________
+5.6655 V