Agilent Technologies E4445A User Manual

Page of 664
108
Chapter 2
Instrument Functions: A - L
Display
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
2.4.4.3
Bottom
Displays the active function in the bottom-left corner of the display.
Key Path:
Display
,
Active Fctn Position
Readback:
Bottom
reads back to line 3 of this key.
State Saved:
Not saved in instrument state, survives power cycle and preset.
Factory Preset: No effect, persistent variable, survives power cycle and preset.
Factory
Default:
Center
History:
Added with firmware revision A.04.00
Remote Command:
:DISPlay:AFUNction:POSition BOTtom
Example:
:DISP:AFUN:POS BOT
2.4.5
Limits
Limit lines can be defined to compare the data to your defined limits and indicate a pass or
fail condition.
Limits
accesses menus that allow you to create, modify, and change the
properties of limit lines. There are two limit lines in the instrument.
Key Path:
Display
State Saved:
Not saved in instrument state.
SCPI Status Bits/
OPC Dependencies: No OPC dependencies.
Factory
Default: