Chapter 2
123
Instrument Functions: A - L
Display
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
2.4.6.2
Clear Title
Allows you to clear a title from the front-panel display. Once cleared, the title cannot be
retrieved.
Key Path:
Display
,
Title
Factory Preset: No title
Remote Command:
There is no equivalent command, but the example below shows how to enter an empty
title.
Example:
DISP:ANN:TITL:DATA “TEST 3”
2.4.7
Preferences
Accesses menu keys which allow you to turn the graticule and annotation on or off.
Key Path:
Display
State Saved:
Saved in instrument state.
Factory Preset
and *RST:
No title
Factory
Default:
n/a
History:
Added with firmware revision A.03.00
2.4.7.1
Graticule (On Off)
Pressing Graticule turns the display graticules
On
or
Off
.
Key Path:
Display
Factory Preset
and *RST:
On
Factory
Default:
On
Terminators:
Enter
History:
Added with firmware revision A.03.00
Remote Command:
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe]?
Example:
DISP:WIND:TRAC:GRAT:GRID OFF