Agilent Technologies E4445A User Manual

Page of 664
Chapter 2
143
Instrument Functions: A - L
File
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
2.5.2.5
Name
Displays the Alpha Editor and enables you to enter a filename. The numeric keypad can
also be used while entering file names. Press
Enter
or
Return
to complete the name entry.
NOTE
Only capital letters (A-Z) and digits (0-9) may appear in file names (8
characters, maximum). Additionally, file names include a 3 digit extension
which is automatically set by the instrument depending on the file type and
format.
Key Path:
File
,
Save
Remote Command:
The file name is entered as part of the directory/path name that is sent with the SCPI
command. See
2.5.2.6
Dir Up
Moves up one subdirectory level within a directory. If your position is in the top level of the
drive already, it moves up to the drive level and the current drive is highlighted (A: or C:).
Key Path:
File
,
Save
Remote Command:
The directory is entered as part of the directory/path name that is sent with the SCPI
command. See
2.5.2.7
Dir Select
Displays the highlighted directory. See
for more information.
Key Path:
File
,
Save
Remote Command:
The directory is entered as part of the directory/path name that is sent with the SCPI
command. See
2.5.3
Load
Displays the menu key that enables you to load instrument-data files from the selected
drive and directory back into the instrument. This function displays the file list box, which
shows the data-entry fields for the file name, type, destination, and path.
The catalog list box is active and can be used for selecting the file information in the
data-entry fields. Only loadable files that match the current type are shown. Placing the