Chapter 2
Instrument Functions: A - L
FREQUENCY / Channel
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
Range:
Remote Command:
[:SENSe]:FREQuency:CENTer <frequency>|UP|DOWN
[:SENSe]:FREQuency:CENTer?
Example:
FREQ:CENT 5 GHZ
sets the center frequency to 5 GHz
FREQ:CENT UP
changes the center frequency to 5.1 GHz if you use
FREQ:CENT:STEP 100 MHz
to set the center frequency step size to 100 MHz
FREQ:CENT?
2.6.2
Start Freq
Sets the frequency at the left side of the graticule and sets the frequency entry mode to
Start/Stop. When the Start/Stop Frequency entry mode is activate, the start and stop
frequency values are displayed below the graticule in place of center frequency and span.
The left and right sides of the graticule correspond to the start and stop frequencies.
When
Start Freq
reaches the upper frequency limit, the stop frequency is set to the highest
available frequency and the start frequency is changed to be less then the stop frequency
by the minimum span (10 Hz).
Center Freq
will be updated to the stop frequency minus
one-half of the span, or to the stop frequency minus 5 Hz, and
Res BW
and
VBW
will be set to
1 Hz.
Key Path:
FREQUENCY
Dependencies/
Couplings:
Start Freq is unavailable when Segmented sweep is on.
State Saved:
Saved in Instrument State
Factory Preset: 10 MHz
Model
Frequency Range
(with Frequency Offset = 0 Hz)
E4440A
–100.0 MHz
to
26.0 GHz
E4443A
–100.0 MHz
to
7.2 GHz
E4445A
–100.0 MHz
to
13.7 GHz
E4446A
–100.0 MHz
to
44.5 GHz
E4448A
–100.0 MHz
to
51.0 GHz