Agilent Technologies E4445A User Manual

Page of 664
186
Chapter 3
Instrument Functions: M - O
Marker
In
st
ru
m
e
nt
Fun
ct
ions:
M
-
O
Marker Trace
1, 2, or 3, or Auto. In
Auto
mode, the analyzer places markers on the
lowest-numbered trace that is in
Clear Write
mode. If no trace is in
Clear Write
mode, it
places the marker on the lowest-numbered trace in
Max Hold
mode. If there are none, then
it uses the lowest-number trace in
Min Hold
mode, then in
View
mode. For example, if trace
1 is in view, and trace 2 is in clear write, any new marker is assigned to trace 2.
Key Path:
Marker
State Saved:
The
Marker Trace
for each marker is saved in instrument state.
Factory Preset: Auto on, Trace 1
Range:
1 to 3
Remote Command:
:CALCulate:MARKer[1]|2|3|4:TRACe 1|2|3
Puts the marker on the specified trace and turns
Auto
OFF for that marker.
:CALCulate:MARKer[1]|2|3|4:TRACe?
The query returns the number of the trace on which the marker currently resides, even if
that marker is in
Auto
mode.
:CALCulate:MARKer[1]|2|3|4:TRACe:AUTO OFF|ON|0|1
Turning
Auto
off sets the
Marker Trace
value to the number of the trace on which the
marker currently resides.
:CALCulate:MARKer[1]|2|3|4:TRACe:AUTO?
The response to the query will be 0 if OFF, 1 if ON.
Example:
CALC:MARK1:TRAC 2
places marker 1 on trace 2.
3.1.8
Readout
This access a menu that enables you to affect how the x-axis information for the selected
marker is displayed in the marker area (top-right of display) and the active function area
of the display. It only affects the readout on the display of the horizontal position
information (for example, frequency).
NOTE
It does not affect the way this information is sent remotely in response to the
CALC:MARK:X?
command.
Key Path:
Marker
State Saved:
In instrument state, for each marker.
Factory Preset: Frequency for non-zero spans and Time for zero spans.
Remote Command:
:CALCulate:MARKer[1]|2|3|4:X:READout FREQuency|TIME|ITIMe|PERiod