Agilent Technologies E4445A User Manual

Page of 664
Chapter 3
193
Instrument Functions: M - O
Marker Fctn
In
st
ru
m
e
nt
Fu
nct
ions:
M
-
O
To guarantee accurate data for noise-like signals, a correction for equivalent noise
bandwidth is made by the analyzer. The
Marker Noise
function accuracy is best when the
detector is set to
Average
or
Sample
, because neither of these detectors will peak-bias the
noise. The trade off between sweep time and variance of the result is best when
Avg/VBW
Type
is set to Power Averaging.
Auto
coupling, therefore, normally chooses the
Average
detector and
Power Averaging
. The Marker Noise function works with all settings of detector
and
Avg/VBW Type
, but using the positive or negative peak detectors does not gives less
accurate measurement results.
Note that the value when the Y-axis units are watts is the square of the value when the
Y-axis units are volts. For example, when the percent ratio with Y-axis units in volts is 20%
(0.2), the percent ratio with Y-axis units in watts will be 4% (0.2
2
= 0.04). When you read
the value out remotely you have to know whether you are in log (dB) or linear (percent),
and if linear, whether volts or watts.
Key Path:
Marker Fctn
Dependencies/
Couplings:
Video
triggering is not available when the detector is
Average
, therefore
marker functions that would set the detector to
Average
, and thus conflict
with video triggering, are not available when the
Video
trigger is
On
.
Remote Command:
See
for the command to select a function.
Example:
CALC:MARK:FUNC NOIS
turns on marker 1 as a noise marker.
CALC:MARK:FUNC?
returns the current setting of marker function for the
marker specified. In this case it returns the string: NOIS.
CALC:MARK:Y?
returns the y-axis value of the
Marker Noise
function for
marker 1 (if
Marker Noise
is ON for marker 1).
3.2.3
Band/Intvl Power
Measures the power in a bandwidth (non-zero span) or time interval (zero span) specified
by the user. If no marker is on, this key activates the delta pair marker mode. If the
detector mode is set to
Auto
, the average detector is selected. If the
Avg/VBW
type is set to
Auto
,
Power Averaging
is selected, other choices of detector and
Avg/VBW
type will usually
cause measurement inaccuracy. The active marker pair indicate the edges of the band.
Only
Delta Pair
and
Span Pair
marker control modes can be used while in this function,
selecting any other mode (for example,
Normal
or
Delta
) turns off this function.
Key Path:
Marker Fctn
Dependencies/
Couplings:
Video
triggering is not available when the detector is
Average
, therefore,
marker functions that would set the detector to
Average
, and thus conflict
with video triggering, are not available when the
Video
trigger is
On
.
Selecting
Band/Intvl Power
when the marker control function is off, normal,