Agilent Technologies E4445A User Manual

Page of 664
210
Chapter 4
Instrument Functions: P - Z
Peak Search
In
st
ru
m
e
nt
Funct
ions
:
P
-
Z
Remote Command:
:CALCulate:MARKer[1]|2|3|4:MAXimum:NEXT
Example:
CALC:MARK2:MAX:NEXT
selects marker 2 and moves it to the next highest
peak.
4.1.2
Next Pk Right
Moves the marker to the next peak to the right of the current marker. The peak must meet
the defined peak excursion and threshold limits. Peaks that are less than 1% of the current
span away from 0 Hz are ignored. If no valid peak is found, an error “No Peak Found” is
displayed. Press
ESC
to clear this message before attempting another search. (Also see the
and
key descriptions.)
Key Path:
Peak Search
State Saved:
Not part of saved state.
Remote Command:
:CALCulate:MARKer[1]|2|3|4:MAXimum:RIGHt
Example:
CALC:MARK2:MAX:RIGH
selects marker 2 and moves it to the next peak to
the right.
4.1.3
Next Pk Left
Moves the marker to the next peak to the left of the current marker. The peak must meet
the defined peak excursion and threshold limits. Peaks that are less than 1% of the current
span away from 0 Hz are ignored. If no valid peak is found, an error “No Peak Found” is
displayed. Press
ESC
to clear this message before attempting another search. (Also see the
and
key descriptions.)
Key Path:
Peak Search
State Saved:
Not part of saved state.
Remote Command:
:CALCulate:MARKer[1]|2|3|4:MAXimum:LEFT
Example:
CALC:MARK2:MAX:LEFT
selects marker 2 and moves it to the next peak to
the left.
4.1.4
Min Search
Moves the active marker to the minimum detected amplitude value on the current trace.