Preset
In
st
ru
m
e
nt
Funct
ions
:
P
-
Z
4.2
Preset
Presetting the instrument provides a known convenient starting point of the instrument
state for making measurements. There are three possible actions when you press the
Preset
key:
For preset type
Mode
(default), the green
Preset
key immediately performs a mode preset.
See the descriptions below. Press
System
,
Power On/Preset
,
Preset Type
to select the preset
type.
For preset type Factory, the green
Preset
key immediately performs a factory preset.
See the descriptions below. Press
System
,
Power On/Preset
,
Preset Type
to select the preset
type.
For preset type User, the green
Preset
key brings up a menu of preset key choices. You
must press one of these keys to initiate an instrument preset.
Pressing
User Preset
resets to the settings/values that you have previously defined as
the
User
preset state using the
Save User Preset
key.
Pressing
Mode Preset
does not change the mode; it only resets the current mode
settings to the factory defaults.
Pressing
Factory Preset
resets the settings for all the modes to the factory defaults.
The factory preset mode is Spectrum Analysis with continuous sweep. If you are not
already in the Spectrum Analysis mode, it switches to that mode.
Pressing
Save User Preset
saves the current user settings.
None of these instrument presets resets “persistent” functions such as GPIB address,
time/date display style, or auto-alignment state to the factory defaults.
See “Restore Sys
Defaults” on page 231.
If
Preset Type
is set to
Factory
in the
System
menu, pressing the
Preset
front-panel key
performs a factory preset, which implements the following:
Resets the analyzer to Spectrum Analyzer mode.
Brings up the
Freq
menu, with no active function.
Sets certain conditions, in all modes, to their default values.
Clears the input and output buffers and all the trace data.
Amplitude-correction factors are turned off, but remain in analyzer memory.
Limit line testing is turned off, but the limit line tables remain in analyzer memory.
The status byte is set to 0.
NOTE
Recalling any state, including the user preset state, will affect the conditions
of more parameters than are affected by a factory preset. For example,
external preamp gain and input impedance correction are not affected by a
factory preset but will be affected by a user preset.
Key Path:
Front-panel key