Agilent Technologies E4445A User Manual

Page of 664
Chapter 4
285
Instrument Functions: P - Z
Trace/View
In
st
ru
m
e
nt
Funct
ions
:
P
-
Z
4.9.3
Clear Write
Erases any data previously stored in the selected trace and continuously displays signals
during the sweep of the analyzer.
Key Path:
Trace/View
Remote Command:
See
.
Example:
TRAC:MODE WRIT
4.9.4
Max Hold
Maintains the maximum level for each trace point of the selected trace (1, 2 or 3), and
updates each trace point if a new maximum level is detected in successive sweeps.
NOTE
Pressing
Restart
, changing the vertical scale (
Amplitude
,
Scale Type
, Log or Lin)
or turning averaging on (
BW/Avg
,
Average (On)
) restarts the held trace.
Key Path:
Trace/View
Remote Command:
See
.
Example:
TRAC:MODE MAXH
4.9.5
Min Hold
Maintains the minimum level for each trace point of the selected trace (1, 2 or 3), and
updates each trace point if a new minimum level is detected in successive sweeps.
NOTE
Pressing
Restart
, changing the vertical scale (
Amplitude
,
Scale Type
, Log or Lin)
or turning averaging on (
BW/Avg
,
Average (On)
) restarts the held trace.
Key Path:
Trace/View
History:
Added with firmware revision A.02.00
Remote Command:
See
.
Example:
TRAC:MODE MINH