Agilent Technologies E4445A User Manual

Page of 664
290
Chapter 4
Instrument Functions: P - Z
Trig
In
st
ru
m
e
nt
Funct
ions
:
P
-
Z
4.10.7
Trig Slope
Controls the trigger polarity. It is positive to trigger on a rising edge and negative to
trigger on a falling edge.
Key Path:
Trig
Dependencies/
Couplings:
Not available for Free Run.
State Saved:
Saved in instrument state.
Factory Preset: Positive (rising edge)
Remote Command:
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
Example:
TRIG:SLOP NEG
4.10.8
Trig Delay
Allows you to control a time delay during which the analyzer will wait to begin a sweep
after receiving an external or line trigger signal. You can use negative delay to pre-trigger
the instrument.
NOTE
Trigger Delay is not available in Free Run, so turning Free Run on turns off
Trigger Delay, but preserves the value of Trigger Delay.
Key Path:
Trig
Dependencies/
Couplings:
This function is not available when Trigger is Free Run or Video.
This function is not available when Gate is on.
State Saved:
Saved in instrument state.
Factory Preset: Off, 1
µs
Range:
–150 ms to +500 ms
History:
Added with firmware revision A.02.00
Remote Command:
:TRIGger[:SEQuence]:DELay <time>
:TRIGger[:SEQuence]:DELay?
:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:DELay:STATe?