Agilent Technologies E4445A User Manual

Page of 664
344
Chapter 5
One-Button Measurement Functions
Mode Setup (Spectrum Analysis Mode)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
Key Path:
Mode Setup
Factory Preset: Off
History:
Added with PSA firmware revision A.04.00
Remote Command:
[:SENSe]:RADio:STANdard:RPARams[:STATe]ON|Off|1|0
[:SENSe]:RADio:STANdard:RPARams[:STATe]?
Example:
RAD:STAN:RPAR Off
5.3.4
Enable All Measurements
Allows you to specify whether all measurements and radio standards are enabled (set to
Yes
) or not (Set to
No
).
NOTE
Enable All Measurements
will be disabled (greyed out) if a measurement is
entered when
Enable All Measurements
is set to
Yes
and the measurement
entered does not support the currently selected Radio Std. “Select a valid
radio standard/measurement combination to enable.” will be displayed.
Key Path:
Mode Setup
Factory Preset: No
History:
Added with PSA firmware revision A.04.00
Remote Command:
[:SENSe]:RADio:STANdard:EAMeas
YES|NO
[:SENSe]:RADio:STANdard:EAMeas
?
Example:
RAD:STAN:EAM YES
5.3.5
Autorange of Power Setting (Remote command only)
This command is the remote version of the
Optimize Ref Level
key. If this function is
applicable to the currently selected measurement, this command sets the reference level
and attenuator to optimum values based on the signal present at the input.
NOTE
There is always a minimum of 6 dB of attenuation set to protect the analyzer
input.
History:
Added with PSA firmware revision A.02.00