Agilent Technologies E4445A User Manual

Page of 664
386
Chapter 5
One-Button Measurement Functions
Display (Complementary Cumulative Distribution Function—CCDF)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
5.11.3
Ref Trace
Press
Ref Trace
to toggle the reference trace display function between
On
and
Off
.
Key Path:
Display
State Saved:
Saved in instrument state.
Factory Preset: Off
Remote Command:
:DISPlay:PSTatistic:REFerence[:STATe] OFF|ON|0|1
:DISPlay:PSTatistic:REFerence[:STATe]?
Example:
PST:REF 1
PST:REF?
5.11.4
Gaussian Trace
Press
Gaussian Line
to toggle the Gaussian line display function between
On
and
Off
.
Key Path:
Display
State Saved:
Saved in instrument state.
Factory Preset: Off
Remote Command:
:DISPlay:PSTatistic:GAUSsian[:STATe] OFF|ON|0|1
:DISPlay:PSTatistic:GAUSsian[:STATe]?
Example:
PST:GAUS 1
PST:GAUS?
5.11.5
Preferences
Displays the menu key that enables you to turn the graticule and annotation on or off.
Key Path:
Display
State Saved:
Saved in instrument state.
Factory Preset: No title
Fundamental
Units:
N/A
History:
Added with firmware revision A.02.00