Agilent Technologies E4445A User Manual

Page of 664
426
Chapter 5
One-Button Measurement Functions
Meas Setup (Intermod (TOI))
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
Example:
TOIN:FREQ:SPAN 20 MHz
TOIN:FREQ:SPAN?
5.18.4
Max Mixer Lvl
Enables you to set the relationship between the highest signal that can be displayed (the
reference level) and the input attenuation. The relationship applies whenever the
Attenuation
is set to
Auto
. The relationship is that the attenuation is given by reference level
minus the max mixer level. For example, as the reference level changes, the input
attuanator changes to ensure that a signal at the reference level does not exceed the
Max
Mixer Lvl
setting.
When the TOI measurement is on, this key controls the maximum mixer level,
independent of the previous setting of
Max Mixer Lvl
located under
Amplitude
. When the TOI
measurement is
Off
, the previous maximum
Max Mixer Lvl
is restored.
Setting
Max Mixer Lvl
to
Auto
sets the maximum mixer level to
−30 dBm.
Key Path:
Meas Setup
Factory Preset: –30.00 dBm/Auto
Range:
History:
Added with PSA firmware revision A.02.00
Remote Command:
[:SENSe]:TOIN:FREQuency:MIXer:RANGe[:UPPer] <number>
[:SENSe]:TOIN:FREQuency:MIXer:RANGe[:UPPer]?
[:SENSe]:TOIN:FREQuency:MIXer:RANGe:AUTO OFF|ON|0|1
[:SENSe]:TOIN:FREQuency:MIXer:RANGe:AUTO?
Example:
TOIN:FREQ:MIX:RANG –10 dBm
TOIN:FREQ:MIX:RANG?
TOIN:FREQ:MIX:RANG:AUTO 1
TOIN:FREQ:MIX:RANG:AUTO?
5.18.5
Optimize Ref Level
Sets the reference level to place the tones near the top of the display. When input
attenuation is set to
Auto
, pressing
Optimize Ref Level
sets the input attenuation according to
the auto rules described in
, based on this reference level. You may want to
trade off analyzer noise and analyzer-induced distortion by manually changing the
attenuation setting from this starting point.