Chapter 5
One-Button Measurement Functions
Meas Setup (Occupied Bandwidth—OBW)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
Remote Command:
[:SENSe]:OBWidth:XDB <dB value>
[:SENSe]:OBWidth:XDB?
Example:
OBW:XDB —50 dB
OBW:XDB?
5.21.7
Optimize Ref Level
Sets the input attenuator to optimize the robustness of the measurement, which is its
freedom from errors due to input compression.
NOTE
There will always be a minimum of 6 dB of attenuation set to protect the
analyzer input.
Key Path:
Meas Setup
State Saved:
Saved in instrument state.
Factory Preset: n/a
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example:
POW:RANG:AUTO ONCE