Agilent Technologies E4445A User Manual

Page of 664
Chapter 2
71
Instrument Functions: A - L
Auto Couple
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
2.1
Auto Couple
Coupled functions are functions that are linked/dependent on other functions. Pressing
Auto Couple
displays some of the most important coupled analyzer functions.
An example of a coupled function is Res BW (resolution bandwidth). When Res BW is set to
auto, the Res BW is automatically set based on the span setting.
Coupled functions are affected depending on how they are coupled. For example, Video BW
is coupled to Res BW, so changing Res BW affects Video BW, but changing Video BW does
not affect Res BW. Changing Video BW puts it in manual (instead of auto-coupled). When
Video BW is set to
Man
it is unaffected by Res BW changes.
When a function is in the
Man
state, a # will appear next to its annotation on the display. If
one or more functions are manually set so that the amplitude or frequency measurement
becomes uncalibrated, “Meas Uncal” appears on the top right side of the graticule.
2.1.1
Auto All
Auto-couples all coupled functions. If
Auto All
is pressed all coupled functions are set to
Auto
.
Setting any auto coupled function to
Man
(manual), uncouples that function without
changing the coupling of other functions.
Coupled functions are functions that are linked. When
Auto All
is pressed, the analyzer
automatically couples all instrument settings that have auto/man choices. In most cases,
these auto settings give the most accurate measurements and optimum dynamic range.
When a function is coupled, it is in the
Auto
state. When it is uncoupled it is in the
Man
state.
Key Path:
Auto Couple
Dependencies/
Couplings:
When segmented sweep is on, this key is not available (greyed out).
When
Auto All
is selected:
Resolution BW
couples to:
Span
and
Span/RBW
Video BW
couples to:
Res BW
and
VBW/RBW
Sweep Time
couples to:
Res BW
;
Video BW
;
Detector
;
Span
and
Center Frequency
CF Step
couples to:
Span
in swept spans, to
Res BW
in zero span
Attenuation
couples to:
Ref Level
;
Ext Amp Gain
;
Atten Step
;
Max Mixer Lvl
; and
Int Preamp
FFT & Sweep
couples to:
Res BW
and
Span
PhNoise Opt
(phase noise optimization) couples to:
Res BW
;
Span
and
FFT &
Sweep
sweep type
Detector
couples to: marker functions;
Avg/VBW Type
;
Average On Off
;
Max
Hold
and
Min Hold
Average Type
couples to: the marker functions;
Detector
and
Scale Type
ADC Dither
couples to:
Sweep Type
;
Span
;
Res BW
;
ADC Ranging
and