Agilent Technologies E4445A User Manual

Page of 664
80
Chapter 2
Instrument Functions: A - L
Auto Couple
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
Remote Command:
Use [:SENSe]:ADC:DITHer[:STATe] AUTO
Example:
ADC:DITH AUTO
2.1.6.2
On
When
ADC Dither
is
On
, the linearity of low-level signals is improved. However, the ADC
dynamic range is reduced to make room for the dither. As a result, the noise floor of the
analyzer is somewhat compromised. So making measurements with ADC dither on gives
you better amplitude linearity, but turning ADC dither off gives you a lower noise floor
(better sensitivity).
Key Path:
Auto Couple
State Saved:
Saved in instrument state
Remote Command:
[:SENSe]:ADC:DITHer[:STATe] ON
Example:
ADC:DITH ON
2.1.6.3
Off
When
ADC Dither
is
Off
, the instrument noise floor is improved. If ADC dither is on the ADC
dynamic range is reduced to make room for the dither. As a result, the noise floor of the
analyzer is somewhat compromised. So making measurements with ADC dither on gives
you improved amplitude linearity. but turning ADC dither off gives you a lower noise floor
(better sensitivity).
Key Path:
Auto Couple
State Saved:
Saved in instrument state
Remote Command:
Use [:SENSe]:ADC:DITHer[:STATe] OFF
Example:
:ADC:DITH OFF
2.1.7
ADC Ranging
Accesses the keys to set the ADC ranging to provide for the best signal to noise ratio
(
Bypass
), or for the best FFT speed (
Autorange
).
Key Path:
Auto Couple
State Saved:
Saved in Instrument State