Fluke 124 User Manual

Page of 83
Specifications
Safety
4
73
Multimeter disturbance:
• 
VDC, VAC, and VAC+DC with STL120 and short ground lead.
• 
OHM, CONT, DIODE, and CAP with STL120, and black test lead to COM.
Table 3
Disturbance less than 1% of full scale
E= 3 V/m
E= 10 V/m
Frequency range 10 kHz to 27 MHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500 
µF
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500 
µF
Frequency range 27 MHz to 1 GHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500 
µF
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500 
µF
Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale.