Warning indications 
 
 
47 
 
Error codes
 
When an error occurs, the three-digit display will indicate the error code. 
 
Display
 
Error name
 
Error description
 
E00 
Interlock emergency stop 
The  interlock  contact  was  opened  during 
irradiation.
 
CH1-4  E10 
Connection error 
The LED head was disconnected.
 
CH1-4  E20 
LED break error 
A  circuit  break  was  detected  in  the  LED 
head during irradiation.
 
CH1-4  E30 
LED short-circuit error 
A short-circuit of the LED head 
was detected during irradiation. 
CH1-4  E40 
LED temperature error 
The  LED  head  temperature  reached  an 
abnormal level.
 
CH1-4  E50 
LED time error 
The  cumulative  time  of  the  LED  has 
reached the set replacement time.
 
CH1-4  E60 
STEP1 calibration error 
CH1-4  E61 
STEP2 calibration error 
CH1-4  E62 
STEP3 calibration error 
CH1-4  E63 
STEP4 calibration error 
CH1-4  E64 
STEP5 calibration error 
CH1-4  E65 
STEP6 calibration error 
CH1-4  E66 
STEP7 calibration error 
CH1-4  E67 
STEP8 calibration error 
CH1-4  E68 
STEP9 calibration error 
CH1-4  E69 
STEP10 calibration error 
The set UV intensity was not reached.