Chapter 6 Ventilator Alarms
109
Alarm
Priority
Audible
Visual (Alarm 
Indicator Button 
and Display)
Device 
Action
User/Clinician Action
Low Circuit 
Leak
High
• • •   • •
Red flashing button; 
“Low Circuit Leak” 
message
Operates
User: Press Alarm Indicator/
Audio Pause button to 
silence alarm. Or, press Left 
(Reset) button to reset alarm. 
Check for occlusions in 
exhalation devices. Make sure 
exhalation device is clean 
and functioning properly. If 
alarm continues, contact your 
health care professional.
Clinician: Remove 
obstruction in leak device. 
If alarm continues, have the 
device serviced. 
High Expiratory 
Pressure
High
• • •   • •
Red flashing button; 
“High Expiratory 
Pressure” message
Operates
User: Press Alarm Indicator/
Audio Pause button to 
silence alarm. Or, press Left 
(Reset) button to reset alarm. 
Make sure tubing is not 
kinked or pinched. Check 
patient’s breath rate. If alarm 
continues, contact your 
health care professional.
Clinician: Check patient’s 
tubing to make sure it is not 
kinked or pinched. If alarm 
continues, have device 
serviced.
Low Expiratory 
Pressure
High
• • •   • •
Red flashing button; 
“Low Expiratory 
Pressure” message
Operates
User: Press Alarm Indicator/
Audio Pause button to silence 
alarm. Or, press Left (Reset) 
button to reset alarm. Make 
sure tubing is not kinked or 
pinched. If alarm continues, 
contact your health care 
professional.
Clinician: Check patient’s 
tubing to make sure it is not 
kinked or pinched. If alarm 
continues, have device 
serviced.