Philips V200 User Manual

Page of 328
Chapter 13
13-40
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Commands 
Transmitted to the 
Ventilator
Commands are transmitted as a series of four ASCII characters followed by a 
carriage return. Valid commands will be stored and response transmissions will 
be sent in the order the corresponding commands were received. 
Transmission of Data 
from the Ventilator
Unless stated otherwise, all fields will be left justified and six (6) characters in 
length. A comma will separate each field, except for the field following the 
start code. Each data transmission shall be terminated with a carriage return.
In the following subparagraphs, a space is designated as “
◆”. When data is 
unavailable, the output field shall contain “
◆◆◆◆◆◆” 
SNDA<CR>, Send 
Variable Length 
Ventilator Settings 
Report
When the ventilator receives SNDA followed by a carriage return, it will 
respond by transmitting the information shown in Table 13-4: “Ventilation 
Report”. The ventilator responds to the SNDA command by returning a string 
with a variable length. The second and third fields define the length of the 
message. The second field indicates the number of characters between the 
start and stop codes. The third field indicates the number of fields between 
the start and stop codes. The fourth field is the start code, 02H. The last field 
in the string is the stop code, 03H, indicating the end of the message. 
NOTE:
The SNDA command is only available in Normal mode.
Ventilation Report
Description
Example
Resolution
Range
Units
Comments
Command Name
MISCA
N/A
N/A
N/A
5 character field
Number of characters 
between the start and 
stop codes
706
N/A
N/A
N/A
3 character field
Number of fields 
between the start and 
stop codes
97
N/A
N/A
N/A
2 character field
Start Code
0x02
N/A
N/A
N/A
ASCII Start 
Transmission 
Character (STX)
Time of request
13:45
N/A
N/A
N/A
24 hour clock, 
hh:mm
Not used
N/A
N/A
N/A
18 character field. 
Outputs example 
value.
Table 13-4: Ventilation Report  (Sheet 1 of 10)