Philips V200 User Manual

Page of 328
Chapter 8
8-8
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Operating Instructions
Front Panel Keys
The front panel keys enable you to initiate ventilator functions. The keys that 
include an indicator (
) also provide operational status of the function that it 
performs. The text version of the front panel keys is shown in Figure 8-7. The 
symbol version of the front panel keys is shown in Figure 8-8. (Refer to 
Chapter 9, “Alarms” for more detailed information on the alarm keys and 
buttons).
Figure 8-7: Front Panel Keys — Text Version
Figure 8-8: Front Panel Keys — Symbol Version
Front Panel Keys
Key Symbol
Definition
Description
Screen 
Lock
Function: Locks and unlocks the graphic display (touch screen). When the 
screen lock is activated, all on-screen buttons are disabled until the touch 
screen is unlocked. This prevents inadvertent setting and display changes 
via the touch screen. MANUAL BREATH, 100% O2, EXP. HOLD, ALARM 
RESET, and ALARM SILENCE keys are still active keys. 
Indicator: Illuminates green when active.
Accept
Function: Enables you to accept selected settings on the front panel 
graphical display.
Table 8-3: Front Panel Keys (Sheet 1 of 3)