Synchronous Operation Mode (AWG710B only)
3-250
AWG710&AWG710B Arbitrary Waveform Generator User Manual
restrictions
There are the following restrictions in synchronous operation mode operation.
„
The same model and the same firmware version should be used for 
Synchronous Operation.
„
When the instruments run in Synchronous Operation mode, you cannot use FG 
mode or Waveform Mixing mode.
„
The Gated mode in Run mode cannot be used. An event jump cannot be used 
although Enhanced mode can be used.
„
Waveform length needs to be in agreement Although waveforms may differ in 
the waveform file loaded to Master and Slave, the number of points of 
waveform files must be the same.
„
The sequence files loaded to Master and Slave needs to be completely the 
same. However, the waveforms specified to be each sequence line may differ, 
as long as the waveform length is the same.
„
The file loaded to the waveform memory cannot be edited or Quick edited.
„
Another waveform file cannot be loaded during output operation (RUN LED 
under lighting).
Connecting the cables
LAN cable
In Synchronous Operation mode, the Master control the Slave through a LAN. 
Master and Slave are connected to a network using a LAN cable. You need to 
connect the LAN cables before you power on the instruments.
Cables for Clock and
Trigger
To synchronize the clock and jump timing on the overall system, connect the two 
types of signals, Sync Clock and Sync Trigger. It is necessary to supply the same 
signals to Master and Slave from Master. The cable to be used must conform to the 
requirements of length and characteristics. Use the specified 50 cm 50 
Ω
 SMA 
cable (P/N: 174-1427-00 made by Tektronix).
Connect the eight SMA cables as shown in Figure 3-86 on page 3-251 . The label 
C of a connection figure expresses the clock and T expresses the trigger signal.