Appendix B: Performance Verification (AWG710B)
AWG710&AWG710B Arbitrary Waveform Generator User Manual
B-19
Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
These procedures check the accuracy of the amplitude and offset outputs of the 
AWG710B Arbitrary Waveform Generator. 
Check Amplitude
Accuracy
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 
Ω 
BNC coaxial cable, a SMA(Ma)–BNC(Fe) adapter, a 50 
Ω
 
precision terminator, and a BNC–to–dual banana adapter to connect the 
AWG710B Arbitrary Waveform Generator CH1 output to the DMM input 
connector (see Figure B-9).
Figure B-9: Amplitude accuracy initial test hookup
NOTE. 
The amplitude and offset accuracy checks are structured as a continuous 
test. The next test uses the control settings from the previous test and uses the next 
step in the sequence file.
NOTE. 
When you output signal from the CH1 or CH1 OUTPUT, check that the 
other OUTPUT (CH1 or CH1) LED is off.
 If the other OUTPUT LED is on, push the CH1 or CH1 OUT button to turn off the 
output.
Equipment 
required
A 50 
Ω
 BNC coaxial cable, a 50 
Ω
 precision terminator, a 
SMA(Ma)–BNC(Fe) adapters, BNC (female)–to–dual banana adapter, 
and a digital multimeter (DMM).
Prerequisites
The AWG710B Arbitrary Waveform Generator must meet the 
prerequisites listed on page B-7.
AWG710B Arbitrary Waveform Generator
DMM
50 
Ω
 BNC coaxial cable
CH1 output connector
+ SMA(Ma)-BNC(Fe) adapter
+ BNC 50 
Ω
 precision terminator
+ 50 
Ω 
BNC coaxial cable
Input connector
+ BNC-to-dual banana adapter
+ 50 
Ω
 BNC coaxial cable