Tektronix AWG710B User Manual

Page of 580
Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual
B-77
Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
These procedures check the accuracy of the AWG710 Arbitrary Waveform 
Generator direct waveform outputs; amplitude and offset. 
Check Amplitude
and
DC Offset
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 
Ω 
BNC coaxial cable, an SMA(Ma)–BNC(Fe) adapter, a 50 
Ω
 
precision terminator, and a BNC–to–dual banana adapter to connect the 
AWG710 Arbitrary Waveform Generator CH1 output to the DMM input 
connector (see Figure B-44).
Figure B-44: Direct DA output amplitude accuracy initial test hookup
2. Set the DMM controls as follows:
3. Push UTILITY (front–panel)!System (bottom)!Factory Reset 
(side)!OK (side).
Equipment 
required
A 50 
Ω
 BNC coaxial cable, a 50 
Ω
 precision terminator, an 
SMA(Ma)–BNC(Fe) adapter, BNC (female)–to–dual banana adapter, and 
a digital multimeter (DMM).
Prerequisites
The AWG710 Arbitrary Waveform Generator must meet the prerequisites 
listed on page B-61.
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 V
Input  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Front
AWG710 Arbitrary Waveform Generator
DMM
50
Ω
 BNC coaxial cable
CH1 output connector
SMA(Male)-BNC(Female) adapter
+ BNC
 50
 
Ω
 precision terminator
50
 
Ω 
BNC coaxial cable
Input connector
BNC-to-dual banana adapter
50 
Ω
 BNC coaxial cable